TY - JOUR
T1 - A system to investigate and adjust profile pattern of computed tomography dose index along the longitudinal-axis
AU - Anam, Choirul
AU - Amilia, Riska
AU - Naufal, Ariij
AU - Dwihapsari, Yanurita
AU - Dougherty, Geoff
N1 - Publisher Copyright:
© 2023, Intelektual Pustaka Media Utama. All rights reserved.
PY - 2023/12
Y1 - 2023/12
N2 - The purpose of this study to develop software to extract and investigate the profiles of the tube current and volume computed tomography dose index (CTDIvol) along the longitudinal axis (z-axis). The tube current and CTDIvol were extracted from the Digital Imaging and Communications in Medicine (DICOM) header of every image along the longitudinal axis. We evaluated the profiles of the tube current and CTDIvol from eight computed tomography (CT) scanners. If the CTDIvol did not fluctuate along the fluctuation of the tube currents, then the system will adjust the CTDIvol with tube currents. It is found that TCM is not always activated. If TCM is activated, the profiles of TCM vary from one scanner to another. The Siemens and Philip scanners have adjusted the CTDIvol profile with tube current, but the Toshiba scanner has not. By developed software, CTDIvol profile of the Toshiba can be easily adjusted. In conclusion, software to investigate the profile pattern of CTDIvol along the longitudinal axis has been successfully developed. The software is easy to use and works quickly. From this study, medical staff must be careful when using the CTDIvol along longitudinal axis contained in each DICOM header.
AB - The purpose of this study to develop software to extract and investigate the profiles of the tube current and volume computed tomography dose index (CTDIvol) along the longitudinal axis (z-axis). The tube current and CTDIvol were extracted from the Digital Imaging and Communications in Medicine (DICOM) header of every image along the longitudinal axis. We evaluated the profiles of the tube current and CTDIvol from eight computed tomography (CT) scanners. If the CTDIvol did not fluctuate along the fluctuation of the tube currents, then the system will adjust the CTDIvol with tube currents. It is found that TCM is not always activated. If TCM is activated, the profiles of TCM vary from one scanner to another. The Siemens and Philip scanners have adjusted the CTDIvol profile with tube current, but the Toshiba scanner has not. By developed software, CTDIvol profile of the Toshiba can be easily adjusted. In conclusion, software to investigate the profile pattern of CTDIvol along the longitudinal axis has been successfully developed. The software is easy to use and works quickly. From this study, medical staff must be careful when using the CTDIvol along longitudinal axis contained in each DICOM header.
KW - CT dose index
KW - Indosect
KW - Patient dose
KW - Size-specific dose estimates
KW - Water equivalent diameter (D)
UR - http://www.scopus.com/inward/record.url?scp=85179705682&partnerID=8YFLogxK
U2 - 10.11591/ijphs.v12i4.23291
DO - 10.11591/ijphs.v12i4.23291
M3 - Article
AN - SCOPUS:85179705682
SN - 2252-8806
VL - 12
SP - 1656
EP - 1662
JO - International Journal of Public Health Science
JF - International Journal of Public Health Science
IS - 4
ER -