Level-dependent IMD cancellation for single-transducer DPOAE measurement using a lookup-table method

  • Irwansyah*
  • , Sho Otsuka
  • , Seiji Nakagawa*
  • *Corresponding author for this work
  • Chiba University
  • Med-Tech Link Center

Research output: Contribution to journal › Article › peer-review

Abstract

Distortion product otoacoustic emissions (DPOAEs) are weak cochlear tones (f3 = 2f1 − f2) evoked by two primary tones (f1, f2) and used clinically to assess cochlear function. In practice, f1 and f2 are typically presented using two independent sound sources. This is because using one transducer to output both tones can generate intermodulation distortion (IMD) at the same frequency (f3) as the DPOAE due to transducer nonlinearity, which makes it difficult to distinguish measurement artifacts from biological emissions. As a result, DPOAE measurement is difficult with conventional earphones, which typically use one playback transducer per ear. Here, we describe a method that enables single-transducer DPOAE measurements by adding a cancellation tone at f3 and tuning its amplitude and phase (via a lock-in amplifier technique) to reduce IMD. Since IMD is level dependent and in-ear stimulus levels vary across ears (due to ear-canal acoustics), the required cancellation parameters are therefore stored in a level-specific lookup table (LUT). The proposed method:•Reduces IMD at the DPOAE frequency to enable single-transducer measurement.•Uses LUT parameters to handle level-dependent IMD.•Was validated on 40 ears against a two-transducer baseline.

Original languageEnglish
Article number104008
JournalMethodsX
Volume17
DOIs
Publication statusPublished - Dec 2026
Externally publishedYes

Keywords

  • Ear-canal sound pressure
  • Intermodulation distortion
  • Lock-in amplifier technique
  • Otoacoustic emissions
  • Probe microphone
  • Single-transducer stimulation

Fingerprint

Dive into the research topics of 'Level-dependent IMD cancellation for single-transducer DPOAE measurement using a lookup-table method'. Together they form a unique fingerprint.