Abstract

Fundamental research about Functional Electrical Stimulation (FES) indicates that FES can be used to rehabilitate motoric system of patient experiencing damage at arrangement of central nerve resulted from cerebrospinal of cord injury (SCI) and also stroke. To generate impulse to human body organ activated by FES is difficult and very complex because of nonlinearity response from neuromuscular system, various response of musculoskeletal system such as having long time delay and fatigue phenomenon (muscle fatigue). Therefore, an open-looped control system is inappropriate to employ. Close-looped control is feasibly appropriate to overcome this constraint in FES appliance to yield accurate impulse generator. For the application of closed-loop control at clinical level, FES system must be designed using portable multichannel device to pleasant the patient. In this research, the realized FES stimulator is designed by employing a highly linear boost converter circuit using fixed frequency and duty cycle ranging from 0% to 25%, because it yields relatively linear and higher output voltage swing than FES controlled by frequency.

Original languageEnglish
Pages (from-to)1974-1982
Number of pages9
JournalJournal of Theoretical and Applied Information Technology
Volume95
Issue number9
Publication statusPublished - 15 May 2017

Keywords

  • Boost converter
  • Closed-loop control
  • Duty cycle
  • FES
  • Functional electrical stimulation