TY - JOUR
T1 - Wire-mesh capacitance tomography for treatment planning system of electro-capacitive cancer therapy
AU - Nismayanti, Anis
AU - Baidillah, Marlin R.
AU - Triwikantoro,
AU - Endarko,
AU - Taruno, Warsito P.
N1 - Publisher Copyright:
© 2021, Penerbit UTM Press. All rights reserved.
PY - 2021/10/20
Y1 - 2021/10/20
N2 - The wire-mesh capacitance tomography (WMCT) has been applied to visualize 2D of the distribution of electric field intensity in the treatment planning system (TPS) of electro-capacitive cancer therapy (ECCT) using human head model. WMCT is proposed in this study to estimate accurately the distribution of electric field intensity which is the main optimum factors of ECCT in order to compensate the inaccuracy of TPS ECCT simulation. The experimental and simulation studies were conducted with wire-mesh sensor consisted of 8×8 wire matrix of copper in human head model using two type of helmet ECCT. The result of electric field value at the intersection wire-mesh have been compared between experimental studies and simulation studies. The electric field average value resulted from ECCT helmet-1 is higher than ECCT helmet-2. The average electric field generated by the ECCT helmet-1 is 1585.72 V/m in an air medium, 97.43 V/m in grey matter and 80.58 V/m in the cancer. While the average electric field generated by the ECCT helmet-2 is 1413.28 V/m in an air medium, 64.20 V/m in grey matter and 52.65 V/m in the cancer. ECCT helmet-1 and helmet-2 result the different of electric field distribution pattern. ECCT helmet-1 is more optimal for used to patient has cancer position in the right and bottom, while ECCT helmet-2 is more optimal for used to patient has cancer position in the top and bottom. Based on this study, the compensation error of the electrical field distribution of simulation study can be applied in order to enhance the accuracy of TPS ECCT. Therefore, it envisages that this study paves way the development of better approaches for general electric field therapy.
AB - The wire-mesh capacitance tomography (WMCT) has been applied to visualize 2D of the distribution of electric field intensity in the treatment planning system (TPS) of electro-capacitive cancer therapy (ECCT) using human head model. WMCT is proposed in this study to estimate accurately the distribution of electric field intensity which is the main optimum factors of ECCT in order to compensate the inaccuracy of TPS ECCT simulation. The experimental and simulation studies were conducted with wire-mesh sensor consisted of 8×8 wire matrix of copper in human head model using two type of helmet ECCT. The result of electric field value at the intersection wire-mesh have been compared between experimental studies and simulation studies. The electric field average value resulted from ECCT helmet-1 is higher than ECCT helmet-2. The average electric field generated by the ECCT helmet-1 is 1585.72 V/m in an air medium, 97.43 V/m in grey matter and 80.58 V/m in the cancer. While the average electric field generated by the ECCT helmet-2 is 1413.28 V/m in an air medium, 64.20 V/m in grey matter and 52.65 V/m in the cancer. ECCT helmet-1 and helmet-2 result the different of electric field distribution pattern. ECCT helmet-1 is more optimal for used to patient has cancer position in the right and bottom, while ECCT helmet-2 is more optimal for used to patient has cancer position in the top and bottom. Based on this study, the compensation error of the electrical field distribution of simulation study can be applied in order to enhance the accuracy of TPS ECCT. Therefore, it envisages that this study paves way the development of better approaches for general electric field therapy.
KW - Electric field intensity distribution
KW - Electro-capacitive cancer therapy (ECCT)
KW - Treatment planning system (TPS)
KW - Wire-mesh capacitive tomography (WMCT)
KW - © 2021 Penerbit UTM Press
UR - http://www.scopus.com/inward/record.url?scp=85120777660&partnerID=8YFLogxK
U2 - 10.11113/JURNALTEKNOLOGI.V83.14362
DO - 10.11113/JURNALTEKNOLOGI.V83.14362
M3 - Article
AN - SCOPUS:85120777660
SN - 0127-9696
VL - 83
SP - 109
EP - 115
JO - Jurnal Teknologi
JF - Jurnal Teknologi
IS - 6
ER -